6. CLEANING & MAINTENANCE
1. Switch off the device and unplug from the power socket.
2. Wipe the exterior of the device with a clean, damp (not wet) cloth and mild dishwashing detergent.
Do not use harsh abrasives or solvents, as these may damage the device.
3. DAILY
Clean chamber and tube.
Remove the breathing tube from the chamber.
Rinse the tube with soapy water. Hang up hose with hose ends pointing to floor to dry.
Remove the chamber by pushing down the finger guard and pulling out the chamber.
Pour out & discard the remaining water.
Clean the chamber with soapy water, rinse and dry.
4. WEEKLY
Thoroughly clean the chamber
Soak the inside of the chamber for 10 minutes in a solution of one part white
vinegar to two parts water.
5. Replace the air filter when it becomes significantly discoloured, at least once every
3 months or after 1000 hours machine running time.
Remove the filter holder from the back of the device and take out the filter.
Replace the old filter with a new filter: Ensure the vertical black line is facing
towards the device.
7. SETUP INSTRUCTIONS
HOW TO SET UP YOUR SYSTEM
1. Remove your device from its packaging.
2. Place the device on a low shelf or on the floor beside your bed,
so that the device is positioned below head height.
3. Chamber Setup (only applicable to integrated CPAP devices)
a. Remove one water chamber from the packaging.
b. Remove the blue caps and discard them (Fig.1).
c. Fill the chamber up to the water filling line with distilled water only
(Fig.2).
Never fill the chamber while it is attached to the device.
When moving your device, ensure that the water chamber is empty.
Device failure due to water damage is not covered by warranty.
d. To attach the water chamber to the device, press down the finger
guard and slide the chamber on (Fig.3) (Fig. 4).
e. When the chamber is fitted correctly, the finger guard will click into
place.
4. One end of the tube has a white plastic connector.
Push this onto the outlet on top of the chamber (Fig.5) or CPAP outlet
(Fig.5a).
5. Connect the other end of the tube to the mask.
6. Plug the device power cord into your household power socket and
into the rear of the device.
7. Switch on the device by pressing the on/off button (See Fig.6 ).
The current humidity setting is displayed.
8. For changing the setting refer to section 8 “Controls and Display”.
Your device is now ready for use.